What Is Dead Space on a 3L Calibration Syringe?
Short answer: Dead space is the volume of air that remains inside a 3L calibration syringe, including its outlet, when the piston is pushed all the way in. It does not change the 3 L the syringe delivers, but some testing systems and lab protocols need the dead space value recorded. When you need it, it is measured during calibration and printed on the calibration label.
Why dead space exists
No syringe can push out every last milliliter of air. The outlet, connector, and the space between the piston and the end cap always hold a small volume. On TLC-manufactured 3L syringes, that volume is about 150 mL and is listed on the calibration sticker.
When you need dead space measured
- Your PFT system or spirometer manufacturer asks you to enter the syringe's dead space value.
- Your hospital or lab quality procedure requires dead space to be documented on the certificate.
- You are replacing a syringe and want the new value on file for your equipment records.
If none of these apply, a standard calibration without dead space is all you need.
How to get a dead space reading
Choose Calibration + Dead Space when you request your RMA. It adds $50 to the standard $160 calibration, for $210 total. The measured dead space is printed on your calibration label and certificate.